Environmental protection
Wet and dust-laden atmospheres
The transmitter is dust and sand tight and protected against
immersion effects as defined by IEC 529 (1989) to IP 67 (IP 68
on request) or by NEMA to 4X or by JIS to C0920
Hazardous atmospheres
With or without output meter/integral display
INTRINSIC SAFETY/EUROPE:
ATEX/BASEEFA approval
• EC-Type Examination Certificate no. BAS 99ATEX 1180
- (HART)
II 1 GDT50°C, EEx ia IIC T6/T5 (-40°C Ta +40°C)
T95°C, EEx ia IIC T4 (-40°C Ta +85°C)
- (FOUNDATION Fieldbus)
II 1 GD T70°C, EEx ia IIC T4 (-40°C Ta +60°C)
• EC-Type Examination Certificate no. BAS 00ATEX 1241
- (PROFIBUS-PA)
II 1 GD T70°C, EEx ia IIB T4 (-40°C Ta +60°C)
TYPE "N"/EUROPE :
ATEX/BASEEFA type examination
• Design compliance by Certificate no. BAS 01ATEX 3380X
- (HART)
II 3 GDT50°C, EEx nL IIC T5 (-40°C Ta +40°C)
T95°C, EEx nL IIC T4 (-40°C Ta +85°C)
- (FOUNDATION Fieldbus)
II 3 GD T70°C, EEx nL IIC T4 (-40°C Ta +60°C)
• Design compliance by Certificate no. BAS 01ATEX 3384X
- (PROFIBUS-PA)
II 3 GD T70°C, EEx nL IIB T4 (-40°C Ta +60°C)
FLAMEPROOF/EUROPE:
ATEX/CESI approval;
• EC-Type Examination Certificate no. CESI 00 ATEX 035
II 1/2 GD T80°C, EEx d IIC T6 (-40°C Ta +70°C)
T95°C, EEx d IIC T5 (-40°C Ta +85°C)
CANADIAN STANDARDS ASSOCIATION
and FACTORY MUTUAL :
- Explosionproof: Class I, Div. 1, Groups A, B, C, D
- Dust ignitionproof : Class II, Div. 1, Groups E, F, G
- Suitable for : Class II, Div. 2, Groups F, G; Class III, Div. 1, 2
- Nonincendive: Class I, Div. 2, Groups A, B, C, D
- Intrinsically safe: Class I, II, III, Div. 1, Groups A, B, C,D,E, F, G
STANDARDS AUSTRALIA (SAA)
TS/WCA Approval (HART only)
• Conformity Certificate no. AUS Ex 3117X
Ex d IIC T5 (Tamb +85°C)/T6 (Tamb +70°C) Class 1 Zone 1;
Ex ia IIC T4 (Tamb +85°C) /T5 (Tamb +55°C) T6 Class 1 Zone 0
Process connections
Flush diaphragm flanged seal (**) :
2in or 3in ANSI 150 to 300 RF
DN50 or DN80 DIN ND 16-40 Form C
Wafer seal (backup flange not supplied)
2in, 3in to ANSI; DN50, DN80 to DIN.
Electrical connections
Two 1/2 NPT or M20x1.5 or PG 13.5 or 1/2 GK threaded
conduit entries, direct on housing; straight or angle Harting
HAN connector and one plug, on request.
Terminal block
HART version
Three terminals for signal/external meter wiring up to 2.5 mm
2
(14 AWG) and three connection points for test and
communication purposes.
Fieldbus versions
Two terminals for signal wiring (bus connection) up to 2.5 mm
2
(14 AWG)
Grounding
Internal and external 6 mm
2
(10 AWG) ground termination
points are provided.
Mounting position
Transmitter can be mounted in any position but with a
maximum ± 1° from vertical.
Electronics housing may be rotated to any position. A positive
stop prevents over travel.
Mass (without options)
9 to 12 kg approx (20 to 27 lb); add 1.5 kg (3.4 lb) for AISI
housing.
Add 650 g (1.5 lb) for packing.
Packing
Carton
Hastelloy is a Cabot Corporation trademark
Galden is a Montefluos trademark
(*) Wetted parts of the transmitter.
(**) Bolts and nuts, gasket and mating flange supplied by Customer