Siemens FUS1010 Bedienungsanleitung Seite 111

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 232
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 110
2-71
1010NFM-3J
Section 2
fx (drive) 29
N (burst length) 5
Ltn in 5.316
Vf max CU
Vs max m/s
Vs min m/s 1355.00
Samples/Cycle 29
HF >-0.120
Siemens Dual Path 4SS10G
Site Setup Data
“Automatic” Adjustment Procedure
z In the [Site Setup Data] Menu, press the <Down Arrow> and scroll to the [HF] menu cell. Press
the <Right Arrow> and a pop-up [Manual] prompt will appear.
z Press the <Up or Down Arrow> to select [Automatic] then press <ENT>.
z The current measured correction value is displayed (see below).
z Press <ENT> again to install this correction value which will now appear next to the [HF] menu
cell.
NOTE: The value shown in the [Automatic] pop-up prompt can not be changed and is for user
information only.
fx (drive) 29
N (burst length) 5
Ltn in 5.316
Vf max CU Adjustment
Vs max m/s >Automatic
Vs min m/s 1355.00
Samples/Cycle 29
HF >-0.000
Siemens Dual Path 4SS10G
Site Setup Data
fx (drive) 29
N (burst length) 5
Ltn in 5.316
Vf max CU >Automatic
Vs max m/s = 0.026
Vs min m/s 1355.00
Samples/Cycle 29
HF >-0.026
Siemens Dual Path 4SS10G
Site Setup Data
z If you decide not to use the [Automatic] selection, press any key other than <ENT> to abort the
operation.
Seitenansicht 110
1 2 ... 106 107 108 109 110 111 112 113 114 115 116 ... 231 232

Kommentare zu diesen Handbüchern

Keine Kommentare