Siemens sitrans LU01 Bedienerhandbuch Seite 80

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 114
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 79
Page 74 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02
Parameters
P821 Spike Filter
Dampens spikes in the echo profile to reduce false readings.
Use P821 if interference spikes are on the long shot Echo Profile display.
P822 Narrow Echo Filter
Filters out echoes of a specific width. Enter width of false echo in msec.
Use this for transducer acoustic beam interference (e.g. ladder rungs). Enter the width of
false echoes (in ms) to be removed from the long shot Echo Profile. When a value is keyed
in, the nearest acceptable value is entered.
P823 Reform Echo
Smoothes jagged peaks in the echo profile. Reforms fragmented echoes into one.
Values
0Off
1*On
Related
P817 Profile Pointer Time
P818 Profile Pointer Distance
P819 Profile Pointer Amplitude
P820 Algorithm
P822 Narrow Echo Filter
P823 Reform Echo
P825 Echo Marker Trigger
Values
0 = OFF (preset)
greater = wider
Related
P817 Profile Pointer Time
P818 Profile Pointer Distance
P819 Profile Pointer Amplitude
P820 Algorithm
P821 Spike Filter
P823 Reform Echo
P825 Echo Marker Trigger
Values
0 = OFF (preset)
greater = wider
Related
P002 Material
P817 Profile Pointer Time
P818 Profile Pointer Distance
P819 Profile Pointer Amplitude
P820 Algorithm
P821 Spike Filter
P822 Narrow Echo Filter
P825 Echo Marker Trigger
Seitenansicht 79
1 2 ... 75 76 77 78 79 80 81 82 83 84 85 ... 113 114

Kommentare zu diesen Handbüchern

Keine Kommentare