SITRANS P measuring instruments for pressure
Transmitters for gage, absolute and differential pressure, flow and level
DS III series for absolute pressure
(from differential pressure series)
2/99
Siemens FI 01 · 2009
2
Rated operating conditions
Degree of protection (to EN 60529) IP65
Process temperature
• Measuring cell with silicone oil filling -40 ... +100 °C (-40 ... +212 °F)
• Measuring cell with inert filling liquid -20 ... +100 °C (-4 ... +212 °F)
• In conjunction with dust explosion protec-
tion
-20 ... +60 °C (-4 ... +140 °F)
Ambient conditions
• Ambient temperature
- Digital indicators -30 ... +85 °C (-22 ... +185 °F)
• Storage temperature -50 ... +85 °C (-58 ... +185 °F)
• Climatic class
- Condensation Permissible
• Electromagnetic compatibility
- Emitted interference and interference im-
munity
To EN 61326 and NAMUR NE 21
Design
Weight (without options) 4.5 kg ( 9.9 lb)
Housing material Poor in copper die-cast aluminium, GD-AlSi12 or stainless steel precision casting, mat. No. 1.4408
Wetted parts materials
• Seal diaphragm Stainless steel, mat. No. 1.4404/316L or Hastelloy C276, mat. No. 2.4819, Monel, mat. No. 2.4360,
tantalum or gold
• Process flanges and sealing screw Stainless steel, mat. No. 1.4408, Hastelloy C4, mat. No. 2.4610 or Monel, mat. No. 2.4360
• O-Ring FPM (Viton) or optionally: PTFE, FEP, FEPM and NBR
Measuring cell filling Silicone oil or inert filling liquid (max. 160 bar (2320 psi a) with oxygen measurement)
Process connection ¼-18 NPT and flange connection to DIN 19213 with mounting thread M10 to DIN 19213 or
7
/
16
-20 UNF to EN 61518
Material of the mounting bracket
• Steel Sheet steel, Mat. No. 1.0330, chrome-plated
• Stainless steel Stainless steel, Mat. No. 1.4301 (SS304)
Power supply U
H
Supplied through bus
Terminal voltage on transmitter 10.5 ... 45 V DC
10.5 ... 30 V DC in intrinsically-safe mode
-
Separate 24 V power supply necessary - No
Bus voltage
• Not Ex - 9 ... 32 V
• With intrinsically-safe operation - 9 ... 24 V
Current consumption
• Basic current (max.) - 12.5 mA
• Startup current basic current - Yes
• Max. current in event of fault - 15.5 mA
Fault disconnection electronics (FDE) avail-
able
-Yes
SITRANS P, DS III series for absolute pressure (from differential pressure series)
HART PROFIBUS PA or FOUNDATION Fieldbus