SD77
3. Rotate the coil on the center shaft so that the leaf springs are perpendicular to the clamping plate and
equidistant from the center shaft.
4. Tighten the clamping plate screw.
5. Remove the shims.
6. Move the coil up and down via the center shaft. Treat the center shaft as though it were a feeler gauge being
used to check a point gap. The coil must not rub against the housing or the pole piece, and the leaf springs must
not buckle or snap.
If rubbing is detected, repeat the alignment procedure.
If a leaf spring snaps to a different curvature, the coil is rotated a little too far to the left or right.
Repeat the alignment procedure.
5.5 FLOAT REPLACEMENT
Removal
Refer to the parts list at the rear of this Instruction, and use the following procedure.
1. Invert the transducer.
2. Remove the bottom plate. Keep the transducer inverted to prevent spilling of the Silicon fluid. If the fluid is
spilled, it must be replaced. Refer to Damping Fluid below.
3. Gently push the float downward and remove its retaining cotter. Lift out the float.
4. Lift the float positioning spring off the center shaft.
Installation
Installation is the reverse of removal. A new gasket is recommended, and the bottom plate must be tightened
securely.
IMPORTANT
All traces of Silicon fluid must be wiped from the bottom plate, both sides of the
gasket, and the bottom surface of the housing. Otherwise, the fluid will seep past the
gasket regardless of how much the screws are tightened.
Check the amount of fluid in the bottom chamber before installing the gasket and plate. Refer to Damping Fluid
below.
Damping Fluid
The bottom chamber contains 43cc. of Dow-Corning DC-200 Silicone fluid, 200 centistokes viscosity. Small bottles
of this fluid (enough for a little more than one complete fill) may be ordered: PN 12334-43.
Check the fluid depth by inverting the transducer, removing the bottom plate, and inserting a 6" scale between the
float and the housing. Allow the scale to bottom in the housing. Fluid depth should be approximately 3/4" (19mm).
IMPORTANT
All traces of Silicon fluid must be wiped from the bottom plate, both sides of the
gasket, and the bottom surface of the housing. Otherwise, the fluid will seep past the
gasket regardless of how much the screws are tightened.
34