Mounting the Analyzer, Continued
Refer to Application Drawing Package, Dimension Diagram Analyzer sheet for drilling information.
The following diagram has typical information for mounting, and as shown in the example pattern
below, drill 12 mounting holes in the shelter (16 if sample system is attached to Maxum II.) If using
an M6 bolt, use a 5.10 mm (.201 inch) drill bit. If using ¼ x 20 bolts, use a No. 7 (.201) drill bit. If
using ¼ x 28 bolts, use a No. 3 (.213) drill bit. Be certain mounting height is convenient for the user.
Protective Earth Ground
Located on back of the
Maxum II (see page 16)
Dimensions
for Air Bath
Oven Models
Only
Figure 2-1: Maxum II Outline Drawing
With a forklift or adequate personnel, lift and align Maxum II mounting holes with those of the house
or shelter wall pre-drilled mounting holes.
Insert the 12 or 16 bolts through Maxum II mounting brackets and into mounting wall stud pre-drilled
holes. Before tightening mounting bolts, be certain unit is leveled both horizontally and vertically.
Securely tighten all bolts to be certain the Maxum II is securely mounted to the wall.
CAUTION Ensure ground wire / protective conductor from external protective earthing system is
connected to ground lug / protective conductor terminal before connection to mains power. Refer
to the drawing above for location of the PE lug.
16