Utility Gas Supply Installation
Description
This section provides the utility gas supply installation requirements for the analyzer.
Carrier Supply Gas
Calibration Standard
Detector Fuel Supply
Compressed Air Supply
Refer to Application Drawing Package, Block Diagram Utility Connections sheet for analyzer
requirements.
WARNING Ensure that personnel, electrical, and local and national codes are considered for
storage, use, and handling of compressed and liquefied gasses in portable cylinders.
Carrier Supply Installation
The carrier gas is specified in the Application Drawing Package on the Block Diagram Utility Connections
sheet. See Technical Specifications on page
11 for requirements for the air supply.
Install two secured carrier gas cylinders in tandem near the analyzer mounting location. This allows
exchange of the cylinders without disrupting the carrier supply. Install carrier lines using new, clean, ¼”
(.635 cm) stainless steel tubing and fittings between the carrier gas cylinders and the analyzer. Do not
use cutting oil on the tubing. Install a check valve to prevent contamination if backpressure should occur.
Consumption
Typical carrier consumption is 180 standard cubic feet per month (5100 liters per month.)
Table 2-1 Regulator Assemblies
Regulated Gas
Manufacturer's
Name and Part
Number
Siemens
Part #
Valve & Fitting
Material
Helium or nitrogen of
ultra high purity
ConCoa
2124301-580
X10014 Stainless Steel
Hydrogen of ultra
high purity
ConCoa
2124-310-350
X10035 Stainless Steel
23