Maxum II Initial Startup Procedures
Important
Before proceeding with these procedures make sure the Maxum II is installed correctly in accordance
with these instructions and local and national codes. Refer to the custom documentation for particular
Maxum II details and/or procedures.
Instructions
Perform all the procedures in this section consecutively starting with inspection of the Electronics
Enclosure.
WARNING When the Electronic Enclosure door is open, voltage dangerous to life exists. These
procedures will involve operation of the unit with the electronics door open, which will require a
“hot work permit” in some locations to ensure that there are no hazards for the personnel
working in the area.
Inspection of the Electronics Enclosure
The following procedure visually checks that the connectors and plugs in
the Electronic Enclosure are properly connected.
WARNING/CAUTION The cable harness connectors and the
chassis plugs associated with the Heater power circuits are
marked with orange identifier tags. Before reconnecting any
connector or plug to a Heater power circuit, ensure that the
orange identifier tag on the connector or plug reads identical to
the orange identifier tag on its mating connector.
The ferrules connected on the gas supply side of the EPCM
manifold are composed of vespel-graphite. To prevent
damage, these ferrules must NOT be over-tightened. Proper
tightness is typically ½ turn past finger-tight.
Ensure that the AC (Mains) power is off to the Maxum II.
Open the electronic enclosure door and inspect all connections to the Power Entry Control
Module (PECM.)
Inspect all SNE/DPM/EPC/SVCM connections (see WARNING/CAUTION above.)
Preparation and Inspection of the Oven Enclosure
The following procedure checks the oven operation.
Ensure that all caps from vents on the sides of the oven (outside) have been removed.
WARNING All vent lines and oven exhaust should be vented outside the shelter. The vents
should be connected to a location where the pressure is atmospheric. Connection of oven or
analyzer vents to a variable pressure source such as a flare header should not be done as
chromatographic performance will be severely affected. Backpressure on the oven vent could
cause the T-rating of the instrument to be exceeded, which might become an ignition source.
32