Description
You can use the menu-guided interface of SIMATIC PDM or HART Communicator to select
the range of parameters to be reset:
1.
Current trim
2. Sensor zero point calibration (position correction)
3. Pressure corrections (zero point calibration and sensor trim)
4. All parameters relevant for the processing of measured values, such as zero point, limit
point, electrical damping, display unit, current trim, zero point calibration (position
correction), sensor trim, measuring speed, interrupt current limits, interrupt setting, overflow
ranges of current.
5. Variable mapper. This performs the following setting:
PV= Pressure, SV= Sensor temp., TV= Electronics temp., QV= Non-linearized pressure
PV Primary variable
SV Secondary variable
TV Tertiary variable
QV Quarternary variable
See also
Factory calibration - FAQ (
http://support.automation.siemens.com/WW/view/en/
10806940/133000)
7.19
Static configuration data
Description
A further menu command in the corresponding operating program allows you to read and also
write a series of sensor-specific material data. With the factory state, this data is pre-allocated
according to the particular device model. These values are not contained in the "Factory
calibration" function; in other words, changes in the device are permanently saved.
List of variable material parameters:
Flange type
Flange material
Remote seal type
Filling medium
O-ring material
Remote seal
Remote seal diaphragm material
Number of remote seals
Sensor filling medium
Operating functions through HART
7.19 Static configuration data
SITRANS P, DS III series with HART
126 Operating Instructions, 09/2012, A5E00047092-08