CALIBRATION AND MAINTENANCE UM344-2
September 1995
6
-
16
Check that all wire connections are tight.
Enclosure threads must be coated with a wet, paste-type, anti-seize compound such as Never-Seez by
Emhart Bostik. Inspect the enclosure O-ring for damage.
It is not recommended that the enclosure’s electronics module compartment be opened for inspection.
6.2.5 Transmitter Calibration
An annual calibration check should be performed to ensure that the transmitter is within specifications.
Refer to Section 6.1 for details.
6.3 TROUBLESHOOTING
This Section provides guidance and procedures to assist in identifying and correcting a malfunctioning
Model 344 Temperature Transmitter. Section 6.2.1 lists needed tools and equipment.
All documentation associated with the transmitter including piping and loop wiring diagrams and
configuration documentation should be available to maintenance personnel to facilitate troubleshooting.
NOTE
The Moore XTC Communicator (MXC) is approved for use in non-
hazardous areas only.
6.3.1 Preliminary Troubleshooting
Symptoms and messages resulting from a transmitter malfunction can expedite the troubleshooting
process. Messages are displayed on an MXC screen and the optional digital meter
The following paragraphs explain the displayed messages and other visual indications that show the status
of the loop or the status of the transmitter.
A. Moore XTC Communicator
Use the MXC's STATUS key to interrogate a transmitter and display a FAILSAFE message caused
by a transmitter failing a self-diagnostic test. This will identify the source of error through an error
message described in Table 6.1 and Section 6.3.2.1, paragraph
A.
Transmitter self-diagnostics are executed by selecting TEST
from the MXC's CALIBRATE/TEST screen. If the test is
failed, a message is displayed. Refer to Section 6.3.2.1, paragraph A for details.
CALIBRATE / TEST
CAL TEST END