UM344-2 ON-LINE AND OFF-LINE OPERATION
August 1995
5
-
17
5.1.3.2 Local Pushbutton Input
Ranging (Controller OFF)
The low and high process temperatures selected as the 0%
and 100% span points may be applied to the transmitter and
stored in the EEPROM using the ZERO/FULLSCALE
pushbuttons. This procedure assumes the transmitter is field
mounted to an operating process. If the transmitter does not
contain an optional Digital Meter to report the process
variable, other instrumentation on the Network must be
available to provide this information.
1. Open the sliding cover on top of the Transmitter's
housing to access the ZERO/FS pushbuttons. See Figure
1-1.
2. Change ZERO range value:
1) Supply new ZERO value.
If the process variable currently being reported by
the Transmitter is not the desired ZERO value, then
the process must be manipulated by an operator to
arrive at the desired value.
2) Activate the Pushbutton Mode
Press and hold for 5 seconds or more the ZERO
pushbutton, then release the pushbutton
3) Enter ZERO Value
Press and release the Zero pushbutton. The new
0% range value is stored in memory. If an optional
Digital Meter is installed, observe that the "PB"
annunciator in the display extinguishes (Normal
Mode).
When a new 0% value is set, the Transmitters 100%
value is automatically shifted to maintain the original
span except as follows:
If the process is out of range, no new ZERO
value is stored.
If the new ZERO value would shift the
FULLSCALE value past the sensor limit, the
new FULLSCALE value is automatically set
to the appropriate sensor limit (except in the
case where this would produce a span that is