Page 192 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01
mmmmm
Parameters
Any combination of the following Scope Displays is available.
Two methods to select Scope Displays:
Scrolling
1. Press READING to display the Auxiliary Function symbol.
2. Press ARROWS to access the desired Reading display symbols.
3. Press ENTER with the desired display symbols displayed.
1/0 Values
Alternatively, a four-digit binary value may be entered, where a 0 turns the associated
signal display OFF, and a 1 turns the display ON.
1110 = PCn_ :
Echo Profile, TVT Curve, and Echo Marker displays ON
Echo Lock Window display OFF
Use the Scope Displays after pressing TRANSDUCER to observe the result of
parameter alterations. (Take several measurements to verify repeatability and overcome
Echo Lock (P711) restrictions.)
P815 Echo Time Filtered
The time (in ms) from the transmission of the pulse, to when it is processed.
Display Symbol
PCnu
Echo Profile
TVT Curve
Echo Marker
Echo Lock Win-
dow
Primary Index Transducer
Values Range: 0.0 to 9999 (view only)
Related P816 Echo Time Raw